Title Comparison and Analysis of Polymer-Functionalized Carbon Nanotubes for Enhancement of the Quantitative Detection of Procalcitonin Levels in Human Plasma
Authors Ariadna Schuck 1, Hyo Eun Kim 1, Minhee Kang2,3, Yong-Sang Kim1

*Minhee Kang minhee.kang@samsung.com
*Yong-Sang Kim yongsang@skku.edu
Institutions 1Department of Electrical and Computer Engineering, Sungkyunkwan University , Suwon 16419 , Republic of Korea
2 Biomedical Engineering Research Center,Smart Healthcare Research Institute, Samsung Medical Center, Sungkyunkwan University School of Medicine , Seoul 06351, Republic of Korea
3 Department of Medical Device Management and Research, SAIHST (Samsung Advanced Institute for Health Sciences & Technology), Sungkyunkwan University , Seoul 06355, Republic of Korea
Abstract Procalcitonin (PCT) is one of the core biomarkers for the body’s systemic infl ammatory response that is expressed uniformly
in several organ tissues. PCT concentration levels rise signifi cantly reaching the highest peak within 48 h after the bacte-
rial infection starts. In this study, we evaluated different polymer-functionalized carbon nanotubes (CNTs) to enhance the
performance of a sensing platform for the rapid detection of PCT over a large physiological range. CNTs were treated with
polyethylene glycol (PEG), polyvinylpyrrolidone (PVP), and poly(L-arginine) P(Arg) to improve the nanostructures while
promoting stability and reducing non-specific interaction. Different morphology analyses were performed to investigate the
properties of the polymer-functionalized CNTs and compared them with the pristine material. Screen-printed carbon elec-
trode (SPCE) sensors were modified with the three proposed nanocomposites and electrical characterization are performed.
Afterward, the PCT antibody (100 ng/ml) is immobilized on the working electrodes before the electrochemical measurements
with the analytes. Differential pulse voltammetry (DPV) responses were measured to quantify the PCT (27.43–20,000 pg/ml)
in buffer solution and human plasma samples. CNT-PVP and CNT-P(Arg) electrodes presented a high correlation between
the current peaks and the concentration of PCT. Subsequently, an interference study was carried out with infl ammatory-
related biomarkers that could cause a false positive response, the variation of the signal current was less than 3%. The SPCE
devices also went through validation with the “gold-standard” laboratory method to verify the agreement with the proposed
method. This study regarding the optimization of the CNTs is just the fi rst part of our fi nal target regarding a multiplexed
point-of-care device using a small volume of the blood sample to provide a rapid diagnostic regarding infl ammatory response.
Keyword Procalcitonin, Inflammation-related marker, Electrochemical sensor, arbon nanotubes
PDF File
# 2010년도 발행분 부터는 Springer 의 BioChip Journal 페이지에서 전문을 열람하실 수 있습니다.
# 학회회원 로그인 후 [ Springer BioChip Journal 열람하기 ] 버튼을 클릭하시면 새창으로 열립니다.
→ 전체 목록 보기

Address : Room 804, The Korea Science Technology Center The first building, 22, Teheran-ro 7 Gil, Gangnam-gu, Seoul, Republic of Korea (06130)
TEL : +82-70-7767-9855 , FAX : +82-2-921-9856 , E-MAIL : biochip@biochip.or.kr Copyright © The Korean BioChip Society. All Rights Reserved.