Info.
|
Vol.6 - No.3 (2012.09.20) |
Title
|
Correlation between blood glucose and hematocrit: A new estimation methodology |
Authors
|
Pawan K. Tiwari1, Hakhyun Nam2& Young-Soo Sohn1 |
Institutions
|
1Department of Biomedical Engineering, College of Medical Science,
Catholic University of Daegu, 13-13 Hayang-ro, Hayang-eup, Gyeongsan-si, Gyeongbuk 712-702, Korea
2Department of Chemistry, Kwangwoon University, 447-1 Wolgye- dong, Nowon-gu, Seoul 139-701, Korea
Correspondence and requests for materials should be addressed to
Y. S. Sohn(sohnys@cu.ac.kr) |
Abstract
|
The estimation of hematocrit in the enzyme and mediator configured blood glucose monitoring electrochemical strips is proposed conceptualizing an innovative data mining methodology. This methodology demonstrate assimilating the peak current(IP), and time(tP) to achieve the peak current of the amperometry; as a coordinate point in the indigenously nomenclated HMM(Hematocrit Monitoring Method) IP-tP Array. The array consists of constructing a matrix(9횞7 matrix) that could be extended to a matrix of appreciable size within the computational limitations. The computational results of hematocrit estimation are precise in the experimental range of 10-50% hematocrit volume. A mathematical model for the transient current emanating from an electrochemical strip is also formulated corroborating our experimental results. |
Keyword
|
Hematocrit, Hematocrit estimation, Datamining, Anemic, Biosensor, Biosignal |
PDF File
|
# 2010년도 발행분 부터는 Springer 의 BioChip Journal 페이지에서 전문을 열람하실 수 있습니다.
# 학회회원 로그인 후 [ Springer BioChip Journal 열람하기 ] 버튼을 클릭하시면 새창으로 열립니다.
|